Abstract:
The module of 'Many-to-all' in 3C-analyzer can be used for
3C-MTS/Capture-C data analysis.
The parameters of
the module 'Viewpoint locking' are set in this window (Figure 2).
This module allows users to quickly get the chromosome positions of
many viewpoints at a time, which depend on the chromosome positions
of the primary enzyme site you select.
- Genome sequences in FASTA: Select reference genome sequences in FASTA format.
- Chromosome: chromosome of the interested genomic region.
- Start of genomic region: the start of base pair of the interested genomic region.
- End of genomic region: the end of base pair of the interested genomic region.
- Length of viewpoints: the length of viewpoints. Higher values would increase the counting number of co-localized regions. But avoid overlapped viewpoints each other if there were multiple viewpoints in a genomic region.
The viewpoints
file known as site_info.csv would be generated by 3C-analyzer showed
as Figure 2.
Parameters of
adapter trimming pipeline are set in this window (Fig. 1).
- 3' end and 5' end adapter sequences: The default adapter is consistent with the Illumina's sequencing kit.
- Adapter matching: 3C-analyzer identify adapter sequences based on 8 exact matching or 12 matching read sequences with at most one mismatched base allowed.
- Minimum query length: Sequences less than the minimum query length after adapter removal will be removed from sequence alignment.
The third-party
aligner (the default is Bowtie2) is required in this module (Figure
3). The options in Bowtie2 integrated into 3C-analyzer are listed
below.
Qualification is
done in this step (Figure 4).
- Reads background: Any read sequences determined by sequencing below the number of reads background is marked as none-detection and to be neglected.
- Level of probability: The probability is determined by the exponential distribution density function.
Writing date: 2014.11.20, 2015.02.06
No comments:
Post a Comment